Hill Rom Resuscitaire Radiant Warmer
The Hill Rom Resuscitaire radiant warmer is an infant warmer
used in NICU – neo-natal wards with oxygen blender, temp sensor,
and suction. The Hill Rom Resuscitaire
The Resucitaire Radiant Warmer is sold bio-med certified, patient
ready and comes with a 1 year warranty.
Hill Rom Resuscitaire Radiant Warmer Price
Refurbished Price $2,295.00 with 1 year parts warranty
Hill RThe Hill-Rom Air-Shields Resuscitaire RW82VHA-1C is a
combined infant radiant warmer and resuscitator for use in the delivery
suite or special care baby unit. The unit evaluated incorporated all
optional features.
The radiant heater can be operated in pre-warm, manual and servo controlled temperature modes, and also has an Apgar timer. The resuscitation
module incorporates a suction circuit, patient and auxiliary gas supply systems,
air/oxygen blender and a gas-powered ventilation system called
‘AutoBreath™’. The unit has a large bassinet with an X-ray tray and
vertical height adjustment. A storage drawer and instrument tray is provided
beneath the bassinet. Additional shelving is also available.
INFANT RADIANT WARMER UNIT
Heating Control Heat is delivered to the mattress area using an overhead, microprocessor controlled, radiant heater, which is started by pressing the rectangular power ON/OFF button. Two control modes, pre-warm and manual, are provided on all
Resuscitaire models. Skin temperature servocontrol (baby) mode was also installed on the device tested. All controls and the alarm message panel are on the heater control panel at the top of the control column (Photo 1). Most control switches are covered with a tactile membrane, and lights indicate which mode is in use. Heater mode controls are selected using the switch on the left of the panel.
Pre-warm mode An automatic cycle initially selects 100% heater power (750 W) for 3 minutes, then 60% power for 12 minutes, and then reduces the heater power to 30%, at which level no alarms will activate.
Resuscitaire Radiant Warmer Features
Vertical Height Adjustment (VHA)
X-ray tray
Resuscitation module with AutoBreathä
Air/oxygen blender
Auxiliary gas supply
Suction
Hill-Rom Air-Shields Resuscitaire RW82VHA-1C Radiant Warmer Operation Manual
Hill Rom Resuscitaire Radiant Warmer Specification
Maximum heater power output: 750W
Power settings: 0 to100% in10% increments
Warmer modes: Pre-warm, Manual Baby (servo-controlled)
Temperature display range: 18 – 43°C
Baby mode temperature range: 34 – 38°C
Heater head rotation: 90° left or right
Mattress size: 52cm ´ 65cm tilt: variable+10° to –10°
Apgar Timer: tones at 1 5 & 10 minutes Features
Hospital Direct Medical Equipment
“New, Used and Surplus Medical and Hospital Equipment”
4901 Morena Blvd. #600
San Diego Ca., 92117
9AM to 5PM Mon-Fri
10AM to 2PM Saturday
San Diego Office Phone: (858) 263-4894
Virginia Phone: (540) 327-7376
USA Email: hospitaldirectmedical@gmail.com
Winchester and Richmond Virginia
1114 Fairfax Pike #212
White Post Virginia 22663
Phone: (540) 327-7376
nfant Neonatal NICU Medical Equipment