Philips Heartstart XL Defibrillator M4735A
The Philips Heartstart XL is one of the most popular defibrillators sold
in the used-refurbished medical equipment market. This is a popular defibrillator for surgery centers, emergency departments and general hospital use. iMedical Equipment and Service sells a wide variety of new, used and refurbished which come bio-med
certified and patient ready.
Building on 50 years of Philips experience, the HeartStart XL+ is designed for Resuscitation and Rapid Response Teams. It contains meaningful innovations that can help you confidently and effectively respond to patients across the hospital. This defibrillator is popular in emergency departments, surgery centers, ambulances
and trauma care units that need a commercial grade defibrillator for providing
vital sign data needed for resuscitation.
Philips Heartstart XL Defibrillator M4735A Prices
Refurbished with SP02, ECG and pacing with 1 year warranty price: $1,695.00
NEW with SP02 and pacing and 1 year warranty price $3,995.00
ORDER: Call us at (323) 861-0262 for California or (844) 990-0460 for Virginia
or email us at info@imedical.solutions
Easy-to-use interface is designed to be clear
Choice of modes for all levels of users
Event Review Summaries simplify data collection
Easy to carry and access accessories to save time
AED mode for any patient
Standard defibrillator design to increase efficiency
Active ready–for-use indicators clearly show status
Patient monitoring measurements for continuity of care
Universal defibrillation pads for adults and children
HeartStart Adult/Child or Infant defibrillation pads are designed for manual or AED mode, pacing, cardioversion, and monitoring. No need for specialty pads for infant/child AED mode.
Specifications
-
Environmental Humidity -
Up to 95 % relative humidity
Atmospheric pressure range -
Operating and storage – 1014 mbar to 572 mbar(0 to 15000 ft; 0 to 4500 m)
Shock(a) -
Operating Half-sine waveform/duration ≤11 ms/acceleration ≥ 15.3 G at 3 shocks per face
Shock(b) -
Non-operating Trapezoidal waveform/acceleration 30G/ velocity change 7.42 m/s +/-10% 1 shock per face
Water/solids ingress resistance -
Meets Ingress Protection level IP21
EMC -
Complies with the requirements of standard EN 60601-1-2:2002
Safety -
Meets UL 60601-1 (1st edition)/EN 60601-2-4:2003/EN 60601-1:1990
Mode of operation -
Continuous
AC Line powered -
100-240 VAC/50 or 60 Hz/1-0.46A/Class I Equipment
Battery powered -
Minimum 14.4 V rechargeable lithium ion VAC/Hz
-
-
Physical dimensions Defibrillator Model HeartStart XL+ Size -
Size 9˝ high x 11.6˝ wide x 10.9˝ deep (23 cm x 29.6 cm x 27.9 cm)
-
-
Battery Type -
Rechargeable, lithium ion; see battery label for capacity information
Dimensions -
1˝ high x 4.5˝ wide x 5.7˝ long (23.6 mm high x 116 mm wide x 146 mm long)
Weight -
Approximately 1.5 lbs (0.68kg)
Capacity -
With a new fully charged battery, at 20 °C (68 °F) one of the following: At least 3 hours of monitoring (ECG and SpO2 monitored continuously and NBP sampled every 15 minutes) followed by 20 full-energy charge/shocks. OR at least two hours of pacing (180ppm at 140mA with 40 msec pulse width) while monitoring (ECG and SpO2 monitored continuously and NBP sampled every 15 minutes) followed by 20 full-energy charge/shocks; OR at least 175 full energy charge/shocks
Charge time device off -
With the temperature between 0-35° C (32-95° F) less than 3 hours to 100% capacity and less than 2 hours to 80% capacity
Life -
Approximately 3 yr
Battery indicators -
Battery gauge on battery, capacity indicator on display, power indicators on front of device; flashing RFU indicator, chirp and Low Battery messages on the display for low battery condition. When a low battery message first appears there is still enough energy for at least 10 minutes of monitoring and 6 maximum energy discharges
Battery storage -
Storing the battery for extended periods at temperatures above 40° C (104° F) reduces battery capacity and degrades battery life
-
-
Physical dimensions Weight -
Weight 14.7 lbs (6.6 kg), includes one battery/one new roll of paper/one pads cable. Incremental weight of external standard paddles and paddle tray is less than 3 lbs (1.3 kg)
Standard operator position -
Within one meter (3 feet) of the device
-
-
Power supply Type -
Rechargeable lithium ion battery; AC power using a protectively grounded outlet
-
-
Display Size -
Approximately 6.5″ (16.5 cm) diagonal viewing area
Resolution -
640 x 480 pixels (VGA) with 32 brightness levels per color
Type -
Approximately 6.5″ (16.5 cm) diagonal viewing area
Sweep speed -
20 mm/s nominal (stationary trace; sweeping erase bar) for ECG and SpO2 mm/s
Wave viewing time -
5.2 sec
-
-
Environmental Operating temperature -
0°C to 45°C (32°F to 113°F)
Storage temperature -
20°C to 70°C (-4°F to 158°F)
-
-
Defibrillator Waveform -
Biphasic Truncated Exponential. Waveform parameters adjusted as a function of patient impedance
Shock delivery -
Via multifunction electrode pads or paddles
Shock series -
Configurable energy escalation in a series
Charge time -
3 seconds to the recommended adult energy level (150 Joules) with a new fully-charged battery installed
-
Less than 5 seconds to the selected energy level (up to 200 Joules) with a new fully charged battery installed
-
The device powers on in manual defibrillation mode ready to deliver shock in less than 8 seconds plus applicable charge time assuming an immediate selection of an energy and initiation of a charge even at 90V AC and after 15 maximum energy discharges
-
The device powers on in AED mode ready to deliver shock in less than 17 seconds plus applicable charge time
Leads sensing -
Leads off sensing and PCI sensing for pads/paddles. Apply 500nA rms (571Hz); 200uA rms (32KHz)
Minimum patient impedance range -
25 ohm (external defibrillation);15 ohm (internal defibrillation) ohm
Maximum patient impedance range -
250 ohm. Actual functional range may exceed these values ohm
-
-
Printer Continuous ECG strip -
The Print button starts and stops the strip. The printer can be configured to be run real time or with a 10-second delay. The strip prints the primary ECG lead and a second wave with event annotations and measurements
Auto printing -
The printer can be configured to automatically print on Mark Events, Charge, Shock, and Alarm
Reports -
The following can be printed: Event Summary (Long or Short)/Vital Signs Trends/Operational Check/Configuration/Status Log/Device Information
Speed -
25 mm/s with an accuracy of +/-5% mm/s
Amplitude accuracy -
5% for offset voltages of +/- 300 mV at 5Hz
Paper size -
50 mm wide x 30 m long
-